510 Series for
Gas Monitoring
GA-360E for
Trace Gas
MU-2000 for
Multi-component
INM-700 In-situ for
NOX/O2
VA-3000 for
Multi-Component

![]()
![]()
Selectable response time.
Selectable outputs: 0-1 VDC, 0-10 VDC, and 4-20 mA
Digital outputs: Range setting information
Digital Alarms: Indicate malfunctions or calibration faliure.
Digital Inputs: Zero Cal, Span Cal, Data Hold.
Range Inputs: 4 inputs BCD code.
Each model configured with one of four analyzers for measurements of CO, CO2 , O2, NOX ,
SO2 , CH4 , C2H4 , and NO2; others upon request.
![]()
The 510 series of general-purpose gas analyzers provide continuous monitoring of concentrations of the specific sample gas. The analyzers can be operated from controls on the front panel or by commands from a remote computer. Measurement results are displayed on the front panel and are available to remote data logging systems through an industry-standard interface.
The 510 series can be used for a wide variety of analyses and tests, such as industrial process control and composition analysis, and fixed-source emissions monitoring. Each model is configured using one of four analyzers: the VIA-510 infrared analyzer, the CLA-510 chemi-NOX analyzer, the MPA-510 O2 analyzer, and the FIA-510 THC analyzer.
A high level of sensitivity is achieved through the use of a dual-beam NDIR, Chemiluminescence, Magneto-pneumatic, and Flame Ionization analysis methods. The NDIR analyzers use an infrared absorption method which offers superior sensitivity, selectivity, and stability. With the Magnetopneumatic analyzer, high-sensitivity measurements can be realized free of zero drift due to the use of N2 as the operation gas. The Chemiluminescence analyzers use a dryer unit equipped with automatic reseneation, so ambient air can be used as the source of O3. It also provides 7 ranges and a range of 1:100 (0-20/50/100/1000/2000ppm), making it easy to handle a wide variety of applications. Finally, the Hydrogen Flame Ionization method also allows highly sensitive and precise continuous measurements. Measurements can be performed in 7 ranges and in a range ratio of 0 to 10/30,000 ppmC, making it possible to satisfy a wide variety of applications. These highly accurate, high performance analyzers are suitable for process monitoring and control.
Measurement Ranges
| Model | Gas | Min. | Max. |
|---|---|---|---|
| VIA-510 | Carbon monoxide (CO) | 0-50 ppm |
0-100% |
| VIA-510 | Carbon dioxide (CO2 ) | 0-50 ppm |
0-100% |
| CLA-510 | Nitric Oxide (NO) | 0-20 ppm |
0-2000ppm |
| VIA-510 | Sulfer dioxide (SO2) | 0-50 ppm |
0-100% |
| VIA-510 | Methane (CH4) | 0-50 ppm |
0-100% |
| VIA-510 | Ethylene (C2H4) | 0-50 ppm |
0-100% |
| VIA-510 | Nitrous Oxide (N2O) | 0-50 ppm |
0-100% |
| MPA-510 | Oxygen (O2) | 0-5% | 0-50% |
| VIA-510 | Propane (C3H8) | 0-50 ppm | 0-100% |
| FIA-510 | Total Hydrocarbon (THC) | 0-10 ppm | 0-3% |
Performance
Lowest detection limit: 1.0 ppm
Repeatability: ± 1% of full-scale
Response time: Selectable
Zero drift: <1% (full scale) per day
Span drift: <2% (full scale) per week