The FIA-510 provides continuous measurement of concentrations of total hydrocarbons (THC). The analyzer uses a hydrogen flame ionization detection method and has few moving parts. It performs accurate measurement for long periods and is highly stable.
The FIA-510 can be used for a wide variety of analyses and tests, such as industrial process control and composition analysis, environment-related atmospheric and fixed-source emissions monitoring, and automobile emission analysis.
The FIA-510 can be operated from controls on the front panel or by commands from a remote computer. Measurement results are displayed on the front panel and are available to remote data logging systems through an industry-standard interface.
The instrument may be calibrated automatically or manually and is packaged to fit into a standard 19-inch rack. It is designed to work with the Horiba ESF-510 sample conditioning unit.
Ranges: 0-50, 100, 200, 500, 1000, 5000, 10,000 and 30,000 ppm C as propane
Lowest detection limit: 1.0 ppm
Response time: <60 seconds
Zero drift: ±1% (full scale) per day
Span drift: ±2% (full scale) per day