Few moving parts - superior reliability.
Sensitive and stable.
Selectable response time.
Multiple ranges available.
Digital I/O and isolated analog output.
![]()
The CLA-510SS provides continuous measurement of concentrations of oxides of nitrogen (NOx).
The analyzer uses the chemiluminescent method and has few moving parts. It performs accurate
measurement for long periods and is highly stable. The detector and sample chamber are
specifically designed to minimize interference and quenching effects.
Standard ranges of 0–20, 50, 100, 200, 500, 1000, and 2000 ppm are provided with 0-2.0 or 0-10
ppm minimum ranges optional.
The CLA-510SS can be used for a wide variety of analyses and tests, such as industrial process
control and composition analysis, environment-related atmospheric and fixed-source emissions
monitoring, and automobile emission analysis.
The CLA-510SS can be operated from controls on the front panel or by commands from a remote
computer. Measurement results are displayed on the front panel and are available to remote data
logging systems through an industry-standard interface.
The instrument may be calibrated automatically or manually and is packaged to fit into a standard
19-inch rack. It is designed to work with the Horiba ES-C510 sample conditioning unit.
![]()
Standard ranges:* 0-20, 50, 100, 200, 500, 1000, and 2000 ppm
Lowest detection limit: 1.0 ppm
Response time: <60 seconds
Zero drift: ±1% (full scale) per day
Span drift: ±2% (full scale) per week
* Optional 0-2.0 or -10 ppm minimum ranges are available.